MENU
EEZ - Envox Experimental Zone
Open hardware/open source that bridge the gap between DIY/hobbyist and professional solutions
Search ...
You are here:
Home
EEZ Bench Box 3 (BB3)
Specification and comparison with the EEZ H24005
Main menu
Welcome to EEZ
Open science
Contact us
Login
Projects
BB3-ULX3S integration
EEZ DIB
EEZ Bench Box 3 (BB3)
Introduction
DIB BP3C backplane
DIB AUX Power Supply
DIB STM32F7 MCU module
DIB DCP405 Power module
DIB DCM220 Power module
DIB DCM224 Power module
DIB SMX46 switch matrix module
DIB PREL6 power relays module
DIB MIO168 Mixed I/O module
DIB MIO168 AFE1
DIB MIO168 AFE2
DIB MIO168 AFE3
DIB MUX14D multiplexer module
Specification and comparison with the EEZ H24005
Measurements
Wire harness
Building instructions
Firmware and scripts
User manual
SCPI reference guide
Videos
Galleries
EEZ H24005 Bench Power Supply
CF-DIC Power pre-regulator
EEZ Studio
WebPublish
Archived
September, 2016
June, 2016
April, 2016
December, 2015
November, 2015
October, 2015
Month
Jan
Feb
Mar
Apr
May
Jun
Jul
Aug
Sep
Oct
Nov
Dec
Year
2015
2016
5
10
15
20
25
30
50
100
All
Filter
Binding posts section programming codes
Details
Category:
PSU
Published: 07 December 2015
Last Updated: 02 July 2016
X – don’t care
Notes
:
The Remote sense (
OUTPut:SENSe
) is forbidden when channels are grouped in serial (SYSTem:GROup:SERial).
Before channel grouping in serial or parallel all outputs has to be switched off (
OUTPut OFF
). New output enable can follow only after 20 ms guard time that is required for relay contacts to stabilize (
debouncing
).
Regardless of built-in hardware protection never turn on Out8 and Out9 concurrently.